(VALB96) 96uL Micro Cuvette with In/ Outlet Steel Tubes, Pathlength 30mm, 2 Windows, Z dim 15mm
$299.00
This flow through cuvette is for absorption applications, with a path length of 30 mm. For aqueous samples only. Use up to 100°C. The price is for 1 piece.
- Type: Bonded, For aqueous samples only!
Description
VALB96 Flow-Through Quartz Cuvette is a precision bonded-type micro flow cell designed for absorption spectroscopy where small sample volume and long optical path are required. With a 30 mm optical path length and a total internal volume of 96 μL, this cuvette provides enhanced sensitivity for trace concentration measurements. The cell integrates stainless steel inlet/outlet tubes for secure and stable liquid connections.
Key Advantages
- Long optical path (30 mm) increases absorbance sensitivity for low-concentration samples.
- Micro-volume chamber (96 μL) minimizes sample consumption while maintaining accurate readings.
- Stainless steel tube connectors ensure chemical stability and easy integration with flow systems.
- Ideal for continuous or automated measurements in absorption spectroscopy setups.
- Compatible with UV/VIS wavelength range from 190 – 2500 nm.
Typical Applications
- Continuous monitoring in biochemical and environmental analysis.
- Low-concentration absorbance measurement and kinetic assays.
- Industrial or research flow systems requiring minimal dead volume.
Bonded Fabrication Notes
- Bonded fabrication cuvettes are joined with optical adhesive, providing strong sealing and stable alignment.
- Best suited for aqueous samples; avoid organic solvents or strong acids/bases that may degrade the bonding layer.
- Designed for precise optical consistency and long-term reliability under standard laboratory use.
Flow Cell Specifications
| Type | Flow Through Cuvette | Optical Path Length | 30 mm |
| Article No. | VALB96 | Volume | 96 μL |
| Measuring Method | Absorption Cells | Connector | Stainless Steel Tubes |
| Number of Windows | 2 | Wavelength Range | UV/VIS (190 – 2500 nm) |
| Height | 35 mm | Aperture (W×H) | ø 2 mm |
| Width | 12.5 mm | Depth (Inside) | 30 mm |
| Depth (Outside) | 32.5 mm | Center Height (Z) | 15 mm |
| Fabrication Method | Bonded* | Package | Cell Only |
Note: Bonded quartz flow cells are suitable only for aqueous samples. Avoid strong organic solvents or acids to prevent adhesive degradation and leakage. Clean with mild detergent or deionized water after each use.
Comparison of Fabrication Types
| 1 | 2 | 3 | 4 | ||
|---|---|---|---|---|---|
| *Molded | *Fused | *Bonded / Glued | |||
| UV vis ( 190- 2500 nm) | Yes | Yes | Yes | ||
| Transmission Matched | On Request | No | No | ||
| Transmission > 80% | Yes | Yes | Yes | ||
| Resistant to Acids and Bases (except hydrofluoric acid) | Yes | Yes | No | ||
| Resistant to Organic Solvents | Yes | Yes | No | ||
| Usable upto 600°C (1112°F) | Yes | Yes | Yes | ||
| Usable upto 1200°C (2192°F) | Yes | No | No | ||
| 0 Variations Reading | On Request | No | No | ||
| Fabrication | Molded | Assembled with Quartz Powder | Assembled with Glue | ||
| Adhesives | No | No | Yes | ||
| Storage | Short Term Storage | Clean After Use | Clean After Use | ||
| Available Material | Fused Quartz Material | ||||
| Note | Molded and fused cuvettes and cells are assembled by direct fusion without the use of adhesives throughout the production process. This guarantees resistance against corrosive chemicals and high temperature. Each quartz cuvette and cell has excellent transmission from 190nm to 2500nm. | ||||