Standard Quartz Cuvettes for High Accuracy Spectrophotometry

R. Mavrodineanu and J. W. Lazar

Accurate knowledge of lightpath and parallelism of cuvettes used in spectrophotometry are indispensable parameters that must be determined when accurate transmittance measurements of liquid materials are considered. A description is given of the design and techniques developed at NBS for production of quartz cuvettes (SRM 932) having a nominal radiation pathlength of 10 mm ± 0.03 mm. For each cuvette, pathlength and parallelism are certified with an uncertainty of ±0.0005 mm. The method and instrumentation used to determine these parameters are also described.

Additional Keyphrases   

  • lightpath
  • instrumental
  • pathlength
  • radiation pathlength
  • variables in analysis

True transmittance values can be obtained only by accurate measuring techniques and by taking into consideration all factors that can affect and distort the data (1). Because transmittance is the ratio of two radiation flux intensities, it is necessary that the photometric scale of the spectrophotometer used be accurate. Other important conditions that must be satisfied are: wavelength accuracy, use of adequate spectral bandpass, use of collimated radiation, freedom from reflections (interreflection), fluorescence, polarization, light scatter, optical interferences, particulate matter. Surface conditions and the temperature at which the measurements are performed must be defined, and the material subjected to measurements must be homogeneous (freedom of strain) and stable.

When liquid samples are measured the pathlength of the radiation passing through the solution must be accurately known. This last condition is particularly important, not only when accurate transmittance measurements are contemplated, but also when the molar absorptivity of a chemical species is to be ascertained. In these cases, an uncertainty of 0.1 mm in a nominal radiation pathlength of 10 mm, for instance, will result in an error of 1% in the absorbance measurements, and, similarly, an uncertainty of 0.01 mm will result in an error of 0.1%.

Here, we describe procedures developed and used in the construction of quartz cuvettes having a nominal radiation pathiength of 10 mm ± 0.03 mm, parallelism between the two transparent plates within 2 zm, plate flatness of 2 to 3 fringes (mercury green line) and parallelism of these plates within 2 tm over the whole length of the plates.

Additional conditions to be met were freedom from fluorescence and strain, and the use of fused silica of optical grade.

Fulfillment of these specifications would practically eliminate the radiation pathlength error when absorbance measurements are performed at an uncertainty level of 0.1%. Such an accuracy seems to be at the limit obtainable by the present state of the art when measurements of absorbance are performed on liquid samples, and include the errors of sample preparation and handling, as well as the instrumental uncertainty.

Experimental Procedures

Several designs were contemplated for constructing quartz cuvettes that would satisfy the rather demanding specifications mentioned above. The design illustrated in Figure 1 was finally selected, and was developed in association with E. P. Muth of the NBS Optical Shop, who also made all the parts required for the cuvettes discussed in this work. The final assembly was performed by E. I. Klein of the NBS Glassblowing Shop.

high accuracy spectrophotometry cuvettes
Fig. 1. Fused silica material and parts used for constructing the cuvettes for high-accuracy spectrophotometry

A rectangular-shaped fused-silica structure 48.0 mm high, 13.5 mm wide, and 300 mm long was used to make the body of the cuvettes. Segments approximately 13 mm long were cut from this material and ground and polished, by conventional optical shop techniques, to produce a rectangular segment 10 mm ± 0.03 mm long. The rectangular edges of each segment were finished to a flatness of 1 to 2 fringes (mercury green line), and at one end a small orifice was provided to permit air expansion during the assembly operation.

The length of every segment was checked for accuracy with a standard measuring instrument, illustrated in Figures 2a and b, capable of determining this length with an accuracy of ±0.5 um.

Two plates made of non-fluorescent fused silica 51 mm long, 16 mm wide, and 2 mm thick were ground and polished in a similar manner to a width of 12.5 mm, a thickness of 1.1 mm with a parallelism of 0.002 mm, and a flatness of about 2 fringes (mercury green line) as shown in Figure 1. These plates were checked for thickness with the instrument illustrated in Figures 2a and b, and were attached to the rectangular body by careful welding with city-gas-oxygen torch by using a needle-shaped flame.

The top of the assembled unit was then cut to produce a cuvette 43 mm high, and a rectangular fusedsilica block of 6 mm was attached to the open end of the cuvette by fusion. This block was provided with a standard tapered opening with a diameter of 10 mm at the top. A standard-taper Teflon stopper was used to provide tight closure of the cuvette (Figure 1). The cuvette was finished by grinding the two side walls and bevelling the edges. Its nominal dimensions are given in Figure 3. The unit was stress-released by proper annealing.

standard instrument for plates thickness measurements
Fig. 2a. Standard instrument for measuring the thickness of plates and length of quartz segments
standard instrument for plates thickness measurements 2a
Fig. 2b. Close-up of Figure 2a

The cuvette was then measured to determine the radiation pathlength and the parallelism of the unit. For this purpose the instrument illustrated in Figures 4a and b was used, all measurements being performed at a temperature of 20 ± 0.05oC. measurements were made as described below.

The cuvettes were lightly clamped in a fixture that was bolted to the table of a Moore No. 3 Measuring Machine1 and were adjusted until the two plates that define the radiation pathlength were vertical within the limits of their parallelism. An electronic lever-type gage, with a resolution of 0.125 um, was mounted on the spindle. To reach the bottom of the cuvette, we fabricated and mounted a longer lever on the gage. Use of the longer lever required that a correction factor be applied to all measurements, which decreased the resolution of the gaging system to 0.312 um.

[1The identification of commercial instruments and products is given only to permit reproduction of the work described in this paper. In no instances does such identification imply recommendation or endorsement by the National Bureau of Standards, nor does it imply that the particular equipment or product is necessarily the best available for the purpose.]

quartz standard cuvette for spectrophotometry
Fig. 3. General shape and nominal dimensions (mm) of NBS quartz standard cuvette for spectrophotometry

As shown in Figure 4b, a combination of gage blocks and end plates were wrung together so that the distance between the inside faces of the end plates was 10.0000 ± 0.0001 mm. The gage blocks were then positioned beneath the center of the spindle and the gage lever was offset from the center of rotation of the spindle so that when the spindle was rotated 180o the lever contacted each end plate, the meter readout of the gaging system was at its null (zero) position.

The machine table was then moved to position the cuvette beneath the spindle. The spindle was lowered until the gage lever was at its first measuring position inside the cuvette. The table was moved until the lever contacted the face of the cuvette and the meter was at its zero position. The spindle was rotated 180o the gage lever contacted the opposite side of the cuvette. The meter reading multiplied by the correction factor plus the nominal 10 mm pathlength gave the true pathlength at that point. Measurements were made approximately every 4 mm by lowering the spindle.

Seventeen quartz cuvettes were made and measured at NBS according to the procedures described in this paper, and the values found are presented in Table 1.

From these data we conclude that only three of the 17 cuvettes, or less than 18%, did not meet the specifications established for acceptance.

The same measurements were performed on 10 quartz cuvettes of commercial origin. The results (Table 2) show that only four of the 10 cuvettes were acceptable by the same standards, a rejection of 60%. It must be noted, however, that nine of 10 cuvettes were within the specifications of the manufacturer, which indicate a tolerance of nominal lightpath of 10 mm ±0.01 mm.

instrument to measure the radiation pathlength of cuvettes
Fig. 4a. General view of the standard instrument used to measure the radiation pathlength of cuvettes
instrument to measure the radiation pathlength of cuvettes close up
Fig. 4b. Close-up of Figure 4a

Applications

As a result of these measurements, it was decided to produce at NBS a number of quartz cuvettes, identical to those described in this work, and to certify them for radiation pathlength and parallelism. These cuvettes would then constitute a Standard Reference Material (SRM) in spectrophotometry, and be made available to the public as SRM 932.

radiation pathlengths
Standard Quartz Cuvettes for High Accuracy Spectrophotometry 10

Each cuvette will have a certificate of calibration, worded as given in Appendix I.

The cuvette was designed and produced at the National Bureau of Standards by special techniques and from nonfluorescent optical-quality fused silica. The transparent windows are attached to the body of the cuvette by direct fusion, and the unit was strainreleased by proper annealing. The overall flatness of the transparent windows is within two fringes (546- nm Hg line), and their parallelism is within 2 jim. The radiation pathlength was measured before and after the assembly of the cuvette by using electronic feeler-gage type instruments capable of a resolution of 5 parts in 106.

SRM 932 is a result of the combined efforts of the Analytical Chemistry Division,Institute for Materials Research, the Optical Physics Division of the Institute for Basic Standards, and the Instrument Shops Division. 

E. P. Muth and E. I. Klein designed and assembled the cuvette, respectively. The radiation pathlength measurements were performed by E. G. Erber. 

SRM 932 was developed to meet the needs expressed by the Standards Committee of the American Association of Clinical Chemists under the chairmanship of George N. Bowers, Jr., M.D., Hartford Hospital, Hartford, Conn. 

This work was initiated and performed with the approval and encouragement of 0. Menis, Chief, Spectrochemical Section, Analytical Chemistry Division.

Appendix I

Natioflal Bureau of Standards

Certificate

Standard Reference Material 932 Quartz Cuvette for Spectrophotometry

R. Mavrodineanuand J. W. Lazar

This Standard Reference Material is intended as a reference source when production of accurate spectrophotometric data on liquid samples is considered. It consists of an all-quartz rectangular cuvette designed to fit the holder of conventional spectrophotometers. It is provided with two optically transparent and parallel windows defining a nominal radiation pathlength of 10 mm ± .03 mm. The inner surfaces of the windows are parallel within 0.002 mm. The pathlength and parallelism are certified with an uncertainty of ±0.0005 mm as determined by measurements at positions equally spaced 2 mm from the bottom to within 6 mm of the stopper top. These measurements are made at 20 oC,d the distance between each position is about 4 mm.

radiation measurement

Table 2. Radiation Pathlength, in mm, Determined for 10 Quartz Cuvettes Made by a Commercial Manufacturera

Identification of positions at which measurements were made, starting from position 1, at 6 mm from top and ending with posItion 10, 2 mm from bottom of cuvette.

commercial cuvette measurements
Standard Quartz Cuvettes for High Accuracy Spectrophotometry 11

The general shape and nominal dimensions of the cuvette are illustrated in the figure.2 

The cuvette must be handled with great care and should be held only by the frosted quartz side windows. When not in use, it should be stored in the container provided for this purpose. Extended exposure to laboratory atmosphere and dusty surroundings is to be avoided. 

The overall direction and coordination of the technical measurments leading to certification were performed under the chairmanship of 0. Menis and J. A. Simpson. 

The technical and support aspects involved in the preparation, certification and issuance of this Standard Reference Material were coordinated through the Office of Standard Reference Materials by T. W. Mears.

J. Paul Cali, Chief

Office of Standard Reference Materials

References

1. Mavrodineanu, R., An accurate spectrophotometer for measuring the transmittance of solid and liquid materials. J. Res. Nat. Bur. Stand. (U.S.), 76A, 405-425 (1972); Clarke, F. J. J., Highaccuracy spectrophotometry, ibid., pp 375-403; Mielenz, K. D., Physical parameters in high-accuracy spectrophotometry, ibid., pp 455-467; Burke, R. W., Deardorff, E. R., and Menis, 0., Liquid absorbance standards, ibid., pp 469-482; Burnett, R. W., Accurate measurement of molar absorptivity. ibid., pp483-489.